Technical drawings
JJ X-ray Products
Explore our innovative products and technologies
Slit Systems
Over the last 25 years, JJ X-Ray has designed a large portfolio of slit systems used at x-ray and neutron facilities all over the world. Our top priority is to give the customer the best solution whether they need the slit system to be standard, modified or customized.

Collimators
All collimators are based on a custom design and are divided into three categories defined by the orientation of the foils.

Optics
JJ X-Ray A/S supplies high-precision refractive and reflective optics for synchrotron applications. Their refractive optics include SCD CRLs and lenses in beryllium, aluminum, and nickel for beam focusing. Reflective optics feature customizable mirror systems suitable for UHV/XHV environments, offering options like bending, cooling, and heating for enhanced stability and performance.

Monochromators
JJ X-Ray A/S provides a range of monochromators designed for precise control of X-ray wavelengths in various research and industrial applications. Monochromators are essential for selecting specific wavelengths from an X-ray beam, improving the quality and resolution of diffraction and spectroscopy measurements.

Diagnostics and Analysis
JJ X-Ray offers a range of cutting-edge products designed to enhance the precision and reliability of X-ray analysis systems. Diagnostic and analytic tools are essential for accurately measuring and analyzing X-ray beams, ensuring optimal performance across a variety of applications.

Positioning Systems
JJ X-Ray offers an extensive selection of positioning systems, designed for exact motion and endurance with a focus on scientific sector. Achieving groundbreaking results hinges on precise control, and our systems deliver reliable, repeatable motions for a variety of applications.

Custom Solutions
When working on complex or custom projects, we always work in close collaboration with the customer to make sure all requirements are met, and projects are executed in a timely manner.

Full Beamlines
Read more
IB-C140-AIR
The IB-C140-Air slit system has primarily found application within the neutron community. The slit system uses the same concepts as the other IB-slits but is to handle the significantly larger and heavier blades.
The standard blade materials are B4C or tungsten carbide. Other blades can be accommodated by JJ X-Ray or the customers.
Categories:
Technical Specifications
Aperture size
Maximum: 140 mm x 140 mm
Minimum: Full overlap
Resolution
1 micron per full step
Repeatability
< 2 micron
Accuracy
± 10 microns over the full range
± 1 micron or better with encoders
Limit switches (end-of-travel)
Included as standard on all motions
Guiding
High precision internal rails and carriages
Electrical connections
Limit switches coupled to 2 motor connectors (SUB-D 15 pins male)
Motors
2-phase stepping motors
Mechanical dimensions
495 mm x 495 mm x 78 mm
434 mm x 434 mm x 78 mm (housing only)
Mechanical connections
M6-threaded holes are on the sides, as shown in the drawing.
Weight
≈ 12.0 kg
Outer surface
Anodized aluminum in color nature
Common options
Blades: Various blade shapes and materials
Motors: Custom high-resolution stepping, including IMS motors
Encoders: Back-axle rotary encoders or internal Renishaw linear encoders (absolute or incremental)
Non-magnetic guides and lead screws and extension of motor housings for large distance
Custom connectors